|
|
|
RPR SCREENING TEST FOR SYPHILIS
For the Determination of RPR in Serum
SUMMARY AND EXPLANATION
The Rapid Plasma Reagin (RPR) TEST is a non-Treponemal Flocculation Test that is
used to detect and quantify reagin, an antibody present in serum or plasma from persons
with syphilis, or with other treponemal diseases. Occasionally individuals with other
diseases or conditions may also be reactive in the non-Treponemal Tests.
PRINCIPLES
Treponeme palidum, the etiologic agent responsible for syphilis produces at least two
kinds of antibodies in human injections. Treponemal Antibody-Absorption (FTA-ABS)
Test, or MHA-TP whereas the reagin antibody is detected by non-treponemal tests such
as the RPR Antigen Card Test , in the presence of the reagin antibody in the reactive
sample, the RPR Antigen preparation will produce flocculation consisting of black
clumps against the white background of the test card. By contrast, non-reactive samples
will yield an even light-grey homogenous suspension.
REAGENT: FOR IN-VITRO DIAGNOSTIC USE
Reagent Set Cat. No. S8000 provides:
RPR CARBON ANTIGEN - (Cat. No. 8001)
REACTIVE INGREDIENTS:
0.003% cardiolipin, 0.020-0.022% lecithin, 0.09% cholesterol, 0.0125M EDTA, 0.01M
Na2HPO4,0.01M KH2PO4, 0.1% Thimersoal, 0.0188% Charcoal, and 10% Choline
Chioride. Human Serum containing 0.1% sodium azide as perservative.
PRECAUTIONS:
This product is for in-vitro diagnostic use only. The bottle dispenser should be thoroughly washed and the needle can be checked by the following procedure:
1. Attach needle to a 2ml syringe.
2. Fill the syringe with antigen and eliminate air bubbles, and
3. Count the number of drops delivered in 0.5ml by holding the needle
in a vertical position.
STORAGE & STABILITY:
When not in use, store reagents and controls at 2-8ºC. DO NOT FREEZE. Prior to use,
allow reagents and controls to warm up to room temperature. The antigen should be
agitated gently to ensure homogeneity before use. Remove only enough antigen from the
glass bottle for the day's testing use.
SPECIMEN COLLECTION
EDTA Plasma and unheated or heated serum maybe used. Specimen should be free of
bacterial contamination and hemolysis. Fresh, uncontaminated serum samples may be
stored at 2-8ºC up to 5 days prior to testing. Otherwise, the serum samples should be kept
frozen. Plasma specimens should be tested within 48 hours after that time the specimen
should be discarded.
PROCEDURE
MATERIALS PROVIDED:
1. Reactive Control
2. Minimal Reactive Control
3. Non-Reactive Control
4. Antigen Delivery System: 3 mL Dropping Bottle and Needle which
will deliver 60+\-2 drops/mL
5. Test Cards
MATERIAL REQUIRED BUT NOT PROVIDED:
1. Mechanical rotator set at 100+\-2 r.p.m. with humidity cover
2. Timing device
3. Automatic pipettes
4. Test Tubes
5. Gloves
6. Light source
ASSAY PROCEDURE:
NOTE: All specimens, control serum samples and reagent should be at room temperature
before use.
Qualitative Card Test
1. Person performing this test should refer to the RESULTS section to become familiar with the expected results before performing test. Otherwise, perform test with the controls supplied to become familiar with the expected results. Dispense 1 drop of each control onto separate circles of the test card and follow STEPS 3 to 5 below.
2. Dispense one drop of serum or plasma sample onto a separate circle on the test card with the disposable stirrer pipettes supplied. Use a fresh stirrer pipette for each sample.
3. Using the flat end of the stirrer pipettes, spread the sample over the entire area of the test circle.
4. Mix the carbon antigen reagent well. Attach needle to the dropping bottle. Squeeze the dropping bottle to release air and draw sufficient reagent into bottle. Discard the first few drops and then dispense 1 drop (17 uL) of the antigen (while holding the bottle in vertical position) to a test circle containing a sample. DO NOT MIX the sample and the antigen.
5. Place the card on an automatic rotator and place a humidity cover over card. Rotate at 100 r.p.m. for 8 minutes. Following rotation, a brief hand rotation and tilting of the (3 to 4 times) should be made to aid in differentiating non-reactive from reactive results. Read results macroscopically in the "wet" state under a high intensity
incandescent lamp.
INTERPRETATION OF RESULTS:
A reactive result is indicated by the presence of large aggregates in the centre or
periphery of the test circle. All specimens showing any degree of reactivity or roughness
should be quantitated ( follow Quantitative Test Procedure). Roughness is sometimes an
indication of a sample with a prozone. Minimal reactive samples are indicated by the
presence of small or fine aggregates. A negative ( non-reactive) result will display a
smooth grey appearance.
Quantitative Card Test:
1. Dispense 1 drop (00.5ml) of specimen using stirrer pipette onto circle 1.
2. Using an automatic 0.05ml pipette (or stirrer pipette), dispense 1 drop of 0.9% saline onto circles to be numbered 2 to 5. DO NOT SPREAD.
3. Using an accurate volumetric pipette, dispense 0.05ml of the test sample onto
circle 2. Insert the tip of the pipette into the resulting mixture and mix them by drawing
the mixture up and down the pipette approximately 8 times. Avoid any bubble formation
and transfer 0.05 ml of the mixed sample to the third circle. Repeat this serial dilution
procedure to circle 5 and discard 0.05ml from the last circle. Circles 1 to 5 now represent
a dilution series as follows.
Circle 1 2 3 4 5
Dilution 1:1 1:2 1:4 1:8 1:16
4. Using the flat end of the stirrer pipette, spread the diluted samples over the entire area of the test circles starting at circle no. 5 (highest dilution). Repeat this spreading procedure to circles 4, 3, 2 and 1.
5. Dispense 1 drop of carbon antigen from the dropping bottle to each circle. DO NOT MIX. Place the card onto the automatic rotator and rotate for 8 minutes.
6. Immediately after 8 minutes of rotation, read results microscopically in the "wet" state under a high intensity incandescent lamp. The titre of the sample is the reciprocal of the highest dilution to show microscopic aggregates. (see diagram in the RESULTS section).
7. If the sample is positive in the 1:50 dilution, the dilution series should be
extended as follows:
a. Prepare a 1:50 dilution of non-reactive serum in 0.09% saline. This is to be used for making 1:32 and higher dilutions of specimens to be tested. Dispense 0.05ml of this diluent solution onto circles numbered 2 to 5.
b. Prepare a 1:16 dilution of test specimen by adding 0.1mL of serum to 1.5mL of 0.9% saline. Mix thoroughly. Dispense 0.05 mL of 1:16 dilution of test specimen onto circles 1 and 2.
c. On circle 2, insert the tip of an automatic 0.05 ml pipette into the resulting mixture (sample and diluent) and mix by drawing the mixture up and down the
pipette approximately 8 times. Avoid any bubble formation. Transfer 0.05 mL of the
mixed sample to the next circle. Repeat the mixing procedure. Continue this serial
dilution to circle no. 5 and discard 0.05 ml from this last circle. Circles 1 to 5 now
represent a dilution series as follows:
Circle 1 2 3 4 5
Dilution 1:16 1:32 1:64 1:128 1:256
d. Proceed with the test procedure described under STEPS 4 and 5 of the Quantitative Card Test.
e. Continue dilutions until an end-point titre is reached.
RESULTS:
1:2 1:4 1:8 1:16 1:32
QUALITY CONTROL:
The reactive, minimal reactive and negative controls have been included with the test kit
to monitor the performance of the reagent. If the expected results have not been
observed, the reagent should not be used.
LIMITATIONS
False positive reactions occur occasionally with the RPR Carbon Antigen Test. Such
reactions sometimes occur in drug abuse and in such diseases as lupus erythematosus,
mononucleousis, leprosy, viral pneumonia and after smallpox vaccinations. Reactive
RPR Test specimens should be subjected to further confirmation test as recommended in
the Manual of tests for Syphilis. As with any other serological testing procedure, the
diagnosis of syphilis should not be made on a single reactive result. Temperature of the
reagents and specimens is critical to test outcome. Plasma specimens over 48 hours old
may give erroneous results.
PERFORMANCE CHARACTERISTICS
This RPR Antigen has been tested according to Centers for Disease Control (CDC) RPR
Card Test Procedure. The RPR Test was compared to the CDC RPR Antigen and to
another commercially available RPR Antigen in a clinical evaluation. Atotal of 205
specimens were used in this study and the overall agreement of the results was 100%. All
109 Negative Specimens gave negative results when tested by all 3 types of RPR
Antigen. The 96 Positive Specimens gave positive results when tested by all 3 types of
RPR Antigen. Only 4 out of the 96 positive specimens differed by 1 dilution difference in
titre when quantitative tests were performed. No specific deviation trend has been
observed.
Specimens Pulse Test CDC Antigen Competitor Test
Positive 96 96 96
Negative 109 109 109
REFERENCES
1. Hunter, E.F., Deacon, W.E. and Meyer, P.E., An improved FTA Test for Syphilis, The Absorption Procedure (FTA-ABS) Public Health Reports 79,410-412, 1964.
2. Manual of Tests for Syphilis, Public Health Service Publication. No. 411, 1969.
3. Manual of Tests for Syphilis, Public Health Service Publication. No. 411, 1990